Maintenance and care
62
Step Action
3
Tap on the following 4 crosshairs with the stylus.
Result:
After all 5 positions have been measured, you are prompted to touch
the screen within 5 seconds to save the new calibration data.
4
If the calibration was successful ...
If you were not satisfied with the
accuracy of the calibration ...
… within 5 seconds, use the stylus to
touch the screen at any point.
… wait 5 seconds without touching the
screen with the stylus.
Result:
The new calibration data is
saved. The software returns to the
normal user interface.
Result:
The new calibration data is
discarded and the previous calibration
data is loaded. After the software
returns to the normal user interface,
the calibration can be restarted.
TSLIB calibration utility
Touch crosshair to calibrate
Press the screen to apply calibration.