IDAX Error Messages
69
© 2009 Megger AB
No
Message
Explanation
high.
342
Invalid frequency
The desired frequency is out of the specified limits.
Possible typing error in the C-file.
343
Cable parameter not found:
The specified parameter was not found in the sample
cable.
344
Cable parameters not found
No cable data was found.
345
"Temination box" parameter
not found:
The specified parameter was not found in the
"Termination Box".
346
"Termination box"
parameters not found
No "Termination Box" data were found.
Output voltage is not
within specified levels.
The applied voltage differs from the desired.
351
"InitMeasureZ" settings:
Not possible to perform InitMeasureZ measurement.
This message is followed by one of 340-342.
352
No "Amplifier type"
selected
High Voltage amplifier not selected.
356
Electrometer Overload
Overload in Electrometer instrument during a
measurement.Measurement point is probably
inaccurate. Message also written in point property.
357
Error reading backplane
serial
Problem accessing instrument backplane.
358
Cannot find the Name
option in file
Syntax error in Calibration or Self test files.
359
Cannot find the Connection
option in file
Syntax error in Calibration or Selftest files.
360
Error in file ... at
line ...
Syntax error in Calibration or Selftest files.
Overvoltage
An overvoltage occurred.
Measured capacitances
don't match
Disagreement between values of capacitance
measured for different configurations
Specimen capacitance below
limit.
Measured capacitance below specified limit.
Specimen capacitance above
limit.
Measured capacitance above specified limit.
Measured DC
current>MaxDCCurrent
Measured DC current exceeds limits set by
MaxDCCurrent variable.
Measured hum
current>MaxHumCurrent
Measured interference (hum) current exceeds limits
set by MaxHumCurrent variable.
Errors associated with D-file
No
Message
Explanation
400
No sweep
Error in D-file syntax.
401
No points in sweep
Error in D-file syntax.
403
Wrong model
Error in D-file syntax.
404
Wrong parameter
Error in D-file syntax.
405
Printer error:
Displays error from printer routines.
406
Illegal value for x-min
Illegal settings for axis values.
407
Illegal value for y-min
Illegal settings for axis values.
408
Error: y-max is less than
y-min
Illegal settings for axis values.
409
Error: x-max is less than
x-min
Illegal settings for axis values.
410
Error: Table not exported
The D-file table was not exported properly.
411
No information to show
The information dialog could not display the
information.
412
Cannot read from A:\
(floppy)
Results cannot read from floppy disk. Possibly
damaged floppy.
413
Cannot read from drive
Results cannot read from specified drive.
Содержание IDAX Series
Страница 1: ...2009 Megger AB IDAX User s Manual...