IDAX User's Manual
32
© 2009 Megger AB
2.1.1.3
File / Open Measurement
One or more measurements can be opened by this command depending on selected item in
Test Browser.
Selected item
Data to be opened
Single data file
All sweeps in selected data file
Measurement Sequence
All sweeps in all data files under selected Measurement Sequence
Object
All sweeps in all data files in all Measurement Sequences under
selected Object
2.1.1.4
File / Close Measurement
One or more measurements can be closed by this command depending on selected item in Test
Browser.
Selected item
Data to be closed
Single data file
All sweeps in selected data file
Measurement Sequence
All sweeps in all data files under selected Measurement Sequence
Object
All sweeps in all data files in all Measurement Sequences under
selected Object
2.1.1.5
File / Remake Measurement
In cases when user is dissatisfied with a certain measurement it is possible to repeat this
measurement with the same settings using Remake Measurement as follows:
1. Connect IDAX to the test object if not connected.
2. Select the measurement file, (D-file) to be repeated (obligatory).
3. Choose Remake Measurement. It will create a Measurement Plan, (C-file) from the selected
data file, i.e. all measurement settings from this data file are copied into the new
Measurement Plan, (C-file).
4. Make corrections in Measurement Plan if needed.
5. Press ON on IDAX front panel to start the measurement.
2.1.1.6
File / Save
Save any changes made to the current template or Measurement Plan.
2.1.1.7
File / Save As
1.
Select template which is to be saved under new name.
2.
Invoke Save As dialog from menu File / Save As...