IDAX User's Manual
68
© 2009 Megger AB
No
Message
Explanation
227
"DraftSpecVersion" not
assigned
"DraftSpecVersion" must be assigned. See User's
Manual for options.
228
"Configuration" not
assigned
"Configuration" must be assigned if using a
Termination Box with relays. See User's Manual for
options.
229
Value not assigned
If cat 'P' is assigned, the field can not be left blank.
Errors related to IDAX 206 system and measurement settings
No
Message
Explanation
301
Wrong argument in
"LinSweep" function
Wrong number/type of argument. See User's Manual
Appendix A.6 for options
302
Wrong argument in
"LogSweep" function
Wrong number/type of argument. See's Manual
Appendix A.6 for options.
303
Settings could not be
saved
User settings could not be saved.
306
No DSP device name
The name of the DSP instrument is not valid, contact
a service supplier.
307
Could not open DSP
The DSP instrument did not connect properly.
308
No geometric capacitance
309
Error initializing DSP.
DSP not present or
inactive.
The DSP instrument did not connect properly.
311
Neither the file LastC.icf
nor specified file could
be found
The C-file is not found in the system. Probably
problem with reading/writing to disc or Windows
registry.
312
Unknown "C-File" command
The specified command does not exist. Most likely a
typing error.
313
Wrong number of arguments
Too few or too many arguments in a command.
314
No "Amplitude"
"Amplitude" must be assigned. See User's Manual
Appendix A.5 for options.
315
No "Frequency"
"Frequency" must be assigned. See User's Manual
Appendix A.5 for options.
319
No measurement to comment
Comments are possible during a measurement or
directly after the measurement.
320
No current measurement
No measurement which has been started.
321
DSP variables wrong
The DSP-variables (e.g. MinTime, MinCycles, etc.) are
wrong.
322
DSP code not found
No or wrong DSP code found in the system.
323
Index does not exist
in NT registry
Problems accessing Windows Registry.
324
Could not interpolate
calibration data
Could not interpolate transfer functions.
325
No voltage source selected
A voltage source must be selected before measuring.
327
Could not get info from
DSP
Problems with the connection to the DSP instrument.
329
"MeasurementPerPoints"
must be greater than zero.
One measurement at each frequency must be
performed.
330
DSP reset failed
Problems with the connection to the DSP instrument
331
"Else" without "if"
Syntax error in C-file.
332
"End" without "if"
Syntax error in C-file.
333
Only variables or
constants are allowed
Only variables or constants are allowed in an if–else
statement.
334
Types does not match
Wrong type of data (e.g. string assigned to an
integer).
339
Error updating "D-File"
Measured data was not properly added to the D-file.
340
Invalid voltage
The desired voltage is out of the specified limits.
Possible typing error in the C-file.
341
Invalid current
The applied settings results in a current that is too