IDAX
23
© 2009 Megger AB
Ch0 and Ch1 and their check boxes
The Ch0 and Ch1 check boxes select the channels that are displayed. Trace 0, white, is
proportional to the sample voltage. Trace 1, red, shows the electrometer output voltage, which
is a function of the input current.
Horizontal control
The oscilloscope time base can be set either on auto scale or manual. On auto scale, the time
base is set from the generator signal from the DSP. In the manual mode, the time base is set by
the Time/div slide bar. The time above the slide bar refers to the time per division, the full
oscilloscope window width consists of ten divisions.
Roll
The Roll check box can be used for slow signals that the oscilloscope may otherwise have
difficulties to accurately lock.
Vertical control
The gain of the input channels can be adjusted with the two vertical slide bars next to the Grid
check box at the oscilloscope window. The voltage range is displayed beside the Ch0 and Ch1
check boxes. The offset of the displayed signals can be adjusted by placing the mouse pointer
inside the oscilloscope window and dragging it up and down while holding the mouse button.
Grid
The Grid check box displays or hides the grid.
1.6.4
Changing Language
The language in which IDAX user interface (IDAX software) is displayed can be changed. The
following languages are available:
§
English
§
Swedish
§
German
§
French
§
Spanish
Language is changed selecting menu Configuration / Change Language in IDAX System
Control, which opens a file opening dialog. The selected language file, *.lng, is loaded and it
will be used after restarting the IDAX program. Each language has its own language file, *.lng,
and all language files are located in the directory D:\IDA\Language files\.