IDAX User's Manual
56
© 2009 Megger AB
recommended to avoid this frequency. Instead, two measurements near the power frequency
are performed and the values at power frequency are obtained by interpolation.
10 kV equivalents
The 10 kV equivalents are values that are calculated from the measured values. Using low
voltage, e.g. 200 Vpeak, will result in a certain current through the sample. A linear insulating
material, such as oil and/or paper, will have the same dissipation factor and capacitance also at
a 10 kV test voltage. Therefore, it is possible to calculate the current and power at 10 kV from
the measured values.
2.1.4.9
View / Text View
Displays text behind the Measurement Plan or Template. In Text View all formatting and
measurement settings are shown for user and editing is enabled. It is also possible to view data
files (D-files), but their editing is disabled.
2.1.4.10
View / Show Measurement Templates
Shows or hides Measurement Templates associated to objects in Test Browser. Hiding templates
saves space in Test Browser, which may be recommended.
2.1.5
Results Configuration Menu
There are commands under this menu which allow user to configure output of measurement
results, display desired measured values in both graphical and text format as well as edit
measurement plan(s).
Menu command
Description
Configure appearance and Template and Measurement Plan
editing tools
Configure appearance of graphical output
Configure Table view
Display values of losses and capacitance at power (network)
frequency (50 or 60 Hz)
Select an external program for data analysis
Display available models for presentation of results
2.1.5.1
Configuration / Information
Editing of Measurement and Object Templates and Measurement Plans in Text View mode can
be facilitated by means of this dialog window.