IDAX Error Messages
67
© 2009 Megger AB
No
Message
Explanation
106
Key could not be deleted
Problem accessing Windows Registry.
107
Redeclaration of variable
A variable is declared two or more times in the C–
file.
108
Error writing to file
The log file from a Calibration or a Self test could
not be created.
109
Incorrect data type
The Data type specified does not exist.
112
Cannot start measurement
Displayed when IDAX is not properly set up in order
to perform a measurement. Usually accompanied by
a more detailed message.
113
Error reading Windows
registry
Problem accessing Windows Registry.
Errors associated with C-file
No
Message
Explanation
201
Not a complex number
Variable is expected to be complex number in "X+Yi"
format.
202
Illegal assignment
Wrong type has been assigned to variable.
203
Unknown directive
Incorrect command. Most likely a typing error.
204
No "Data type" specified
Type must be specified before data field. Example:
$
Type = 'R'
205
No "Category" specified
Category must be specified before data field.
Example
$ Cat = 'C'
206
Incorrect "Category"
Wrong variable in category (Cat) has been used.
208
Error on line
Indicates the number of line in C-file, where error
was detected.
209
"VoltageSource" incorrect
"VoltageSource" has to be of a defined type. Most
likely a typing error.
210
Boolean is expected for
check box ("CB")
Check boxes can have only Boolean values.
211
Not an integer
Parameter values must be of the same type as in the
parameter declaration.
212
Not a real
Parameter values must be of the same type as in the
parameter declaration.
213
Not a Boolean
Parameter values must be of the same type as in the
parameter declaration.
214
Unknown "Data type"
"Data type" must be of a defined type. Most likely a
typing error.
216
Wrong number of arguments
Too few or too many arguments in a variable.
217
Syntax error
Undefined error in the syntax. Usually accompanied
by additional information.
218
To few radio buttons in
group ("RB")
There must be at least two buttons in a radio group
219
Not a radio buttons item
("RB")
The function is assigned to a value that does not
match any radio button. Most likely a typing error.
220
Not an array of integers
Parameter values have to be of the same type as in
the parameter declaration.
221
Invalid bitmap "Image"
Image is not a *.bmp file or the file is corrupted.
222
Variable not declared
All new variables must be declared with a "New"
function.
223
"AmplitudeScale" not
assigned
"AmplitudeScale" must be assigned. See User's
Manual for options.
224
"VoltageSource" not
assigned
"VoltageSource" must be assigned. See User's
Manual for options.
225
"FileType" not assigned
"FileType" must be assigned. See User's Manual for
options.
226
"SpecVersion" not assigned
"SpecVersion" must be assigned. See User's Manual
for options
Содержание IDAX Series
Страница 1: ...2009 Megger AB IDAX User s Manual...