background image

www.megger.com

DLRO10 and DLRO10X

6

2.  GENERAL DESCRIPTION

The DUCTER DLRO10 and DUCTER DLRO10X make up a family of low resistance 
ohmmeters that measure resistances in the range from 0.1 μΩ to 2 kΩ. Both 
instruments provide a maximum test current of 10 Amps d.c. which is automatically 
selected according to the value of resistance being tested. The DLRO10X allows you to 
override this automatic selection and select your own test current.

The measured value of resistance is output to the instrument display with indication of 
the units, μΩ, mΩ or Ω. The DLRO10 uses a large LED display and indicators to display 
the value and units respectively, while the DLRO10X contains all the information on a 
clear colour display.

The use of a four terminal measurement technique removes the test lead resistance 
from the measured value and auto current reversal eliminates the effect of standing 
voltages across the test sample. Additionally, before and during a test, test lead 
contact is monitored to further reduce the chance of erroneous readings.

A resistance measurement takes approximately 2.5 seconds and comprises a 
measurement with forward current, reverse current and a display of the average. On 
the DLRO10X all three values are displayed, and optionally, the measurement may be 
limited to forward current only.

The instrument is powered by a rechargeable Lithium ion (Li-ion) battery, which 
permits thirteen hundred 10 Amp tests to be completed on a single charge. The 
battery module contains intelligent circuitry, which prevents damage to the battery 
from overcharging, and a battery state indicator. A removable lid that is hinged at 
the back and held closed by magnetic catches covers the top panel. This lid may be 
removed if required by opening fully and snapping out of its hinges.

Both instruments are protected against accidental connection to external voltages up 
to 600 V d.c. applied between any pair of the four terminals for up to 10 seconds.

Please note that applying 600 V to the potential terminals will cause some 
internal heating. Expect errors of up to 30 digits immediately after the 
application of such voltage. Full accuracy will be restored within two to three 
minutes. 

Both instruments are supplied complete with a pair of DH4 Duplex Handspikes with 
1.2 m leads. Other lengths and terminations are available if required.

3.  General Operation - DLRO10

1

2

3

4

5

6

7

8

Item

Description

Item

Description

1

Test current

5

Excess noise

2

Contact fail indicator

6

Mode control

3

Voltage terminals

7

Brightness

4

Current flowing

8

On/Off

Press the On/Off switch to turn DLRO10 on. All lamps will light, V and I lamps will 
flash, the software version will be displayed and the Ω lamp will light. If the test leads 
have not been connected to the test sample the ‘C’ and ‘P’ lamps will also light. Press 
the On/Off button again to turn the instrument off. If the instrument is not used for 5 
minutes it will be turned off automatically.

3.1   Adjust the brightness control to a comfortable level.

Select the Test mode by pressing the Mode button repeatedly. The indicator lamps will 
cycle through the various Test modes in turn (see the section on Test modes). Press the 
Test button to start a test.

3.2   Test Current Indicators

The test current is selected automatically by DLRO10 and is indicated by the lamps on 
the left of the panel. The measured value is displayed in the main window and the 
units (µΩ, mΩ or Ω) are shown by the lamps to the right of the window.

Содержание DLRO-10

Страница 1: ...DLRO10 and DLRO 10X Digital Low Resistance Ohmmeter User Guide User Guide DLRO10 and DLRO10X Digital Low Resistance Ohmmeter...

Страница 2: ...ltage Warning 8 4 4 Discharge Current Warning 8 4 5 The Menu 8 4 6 Test Menu 9 4 7 Options Menu 9 4 7 1 Retrieve 9 4 7 2 Download 10 4 8 Passbands 10 4 9 Set Clock 11 4 10 Delete Data 11 4 11 Storage...

Страница 3: ...8 7 3 1 Standard Charging 18 7 3 2 WARNING 18 8 SPECIFICATIONS 19 9 TROUBLE SHOOTING 21 10 ACCESSORIES 22 11 REPAIR AND WARRANTY 24 11 1 Instrument Repair and Spare Parts 24 11 2 Approved Repair Compa...

Страница 4: ...to achieve this For testing large inductive resistances the current carrying leads should be connected securely to the item under test before starting the test It is not recommended that duplex handsp...

Страница 5: ...ectrical outlets CAT IV Measurement category IV Equipment connected between the origin of the low voltage mains supply and the distribution panel 1 1 Symbols used on the instrument G Caution Refer to...

Страница 6: ...at the back and held closed by magnetic catches covers the top panel This lid may be removed if required by opening fully and snapping out of its hinges Both instruments are protected against accident...

Страница 7: ...he lamp will not operate if the instrument is turned off The I lamp will light if a current greater than 1 mA is still flowing after a test is completed This suggests that an inductive load has been t...

Страница 8: ...measurement cannot be made if any of these messages is visible on the display 4 3 3 External Voltage Warning If external voltage is applied to the terminals a message EXT VOLTS will flash on the displ...

Страница 9: ...mode 4 7 Options Menu The Options menu has five options that are not related to each other as shown below There is a quick access menu that can be accessed using the numbers keys as follows 0 Retriev...

Страница 10: ...on allows you to set upper and lower limits between which the test result average must fall if it is to be signalled as a Pass GREEN Readings outside these limits will be signalled as a Fail RED Set t...

Страница 11: ...o clear DLRO10X s memory of stored data In case you have selected this Option by accident you will be asked to confirm that you wish to delete the data the default is NO Change this to YES and press E...

Страница 12: ...is automatically limited to 1A When Battery level is low 10 the maximum test current is automatically reduced by one level and will be further reduced when a test has to be abandoned because the batt...

Страница 13: ...he test number to 1 4 15 Keypad Operation The 12 key keypad is similar to those used on mobile telephones in that each key is capable of producing one of several characters depending on the number of...

Страница 14: ...ts on loads that exhibit an inductive component using the test mode indicated by the inductor symbol on the DLRO10 or by using the inductive mode on the DLRO10X When measuring inductive loads it is es...

Страница 15: ...as shown in the next section Testing Using Duplex Handspikes or Individual Leads One of the DH4 handspikes is fitted with two lamps marked L1 and L2 and an extra lead This lead should be plugged into...

Страница 16: ...two measurements Both measurements are displayed by DLRO 10X along with the average while the DLRO 10 displays the average only If the combined resistance of the current leads and test sample is grea...

Страница 17: ...e retaining clips and pull the top of the clip away from the instrument body The module now unplugs from the base of the instrument Plug in the charger or connect to a 12 volt vehicle type battery usi...

Страница 18: ...Charging The battery module is charging at its standard rate Fast Battery state flashing during charging indicates temperature fault 7 3 2 WARNING Connecting to greater than 15 volts can cause perman...

Страница 19: ...e and Safety LED Indication Measurement Mode Manual Auto Continuous Inductive Manual Auto Continuous Inductive Unidirectional Control Fully Automatic Fully Automatic with manual override of maximum cu...

Страница 20: ...to 122 F at reduced accuracy Storage 30 C to 70 C 22 F to 158 F Calibration 20 C 68 F Coefficient 0 01 per C from 5 C to 40 C 0 006 per F from 41 F to 104 F Standard charging 0 C to 40 C 32 F to 104...

Страница 21: ...RR 114 Checksum Failure in EEPROM Calibration constants have been lost The DLRO will continue to work but the accuracy can now not be guaranteed At switch on the display will show the software version...

Страница 22: ...3 172 DLRO10LPU UK Mains Line Power Unit UK plug 1003 093 DLRO10LPU US Mains Line Power Unit US plug 1003 171 DLRO10LPU line power unit The DLRO10 and DLRO10X may be also powered from an optional main...

Страница 23: ...06 18 9 m 30 ft 242006 30 Single Leads Single handspikes 2 for potential measurement 2 m 7 ft 242021 7 5 5 m 18 ft 242021 18 9 m 30 ft 242021 30 Current clips 2 for current connections 2 m 7 ft 242041...

Страница 24: ...r Repair If returning an instrument to the manufacturer for repair it should be sent freight pre paid to the appropriate address A copy of the Invoice and of the packing note should be sent simultaneo...

Страница 25: ...cell is located behind the display Dispose of in accordance with local regulations The lithium ion is classified as an industrial battery For disposal in the UK contact Megger Ltd For disposal of batt...

Страница 26: ...USA T 800 723 2861 USA only T 1 214 333 3201 F 1 214 331 7399 E USsales megger com Megger AB Rinkebyv gen 19 Box 724 SE 182 17 DANDERYD T 46 08 510 195 00 E seinfo megger com Megger USA Fort Collins...

Отзывы: