background image

www.megger.com

DLRO10 and DLRO10X

14

5.  TEST MODES

DLRO10 has 4 test modes which are selected by repeated presses of the Mode button. 
At the end of each test, DLRO10 will display the average of the values obtained with 
forward and reverse current.

DLRO10X has 5 modes. On DLRO10X these modes are selected from the TEST menu 
using the cursor control and Enter key. At the end of a test, DLRO10X will show three 
resistance values; that obtained with forward current, that obtained with reverse 
current and the average value.

5.1   Normal mode

Normal mode makes a single measurement of the resistance of the sample using 
forward and reverse current

Please note that in this mode both the current and voltage leads must be connected 
across the test sample before the Test button is pressed. To make another 
measurement, ensure the test leads are connected and press the test button.

5.2   Automatic mode

If Automatic mode is selected, your instrument will remain in a state of readiness 
waiting for both the current and voltage measuring leads to be connected to the 
test piece. When this is done, a forward and reverse current test will be started 
automatically.

To make another measurement simply break contact with the test sample and remake 
contact.

For example, if measuring joints in a long bus bar, you may leave the current circuit 
connected at opposite ends of the bus bar. Simply making contact with voltage 
probes across the joint(s) you wish to measure will then activate the measurement.

5.3   Continuous mode

Continuous mode requires the connections to be made before pressing the Test 
button. Your instrument will then repeat its forward and reverse measurements and 
display the results approximately every 3 seconds until contact is broken or the Test 
button is pressed.

5.4   Inductive mode

DLRO10 and DLRO10X are able to make d.c. resistive measurements on loads that 
exhibit an inductive component using the test mode indicated by the inductor symbol 
on the DLRO10, or by using the inductive mode on the DLRO10X.

When measuring inductive loads it is essential that the current carrying leads are 
securely clamped to the item being tested and that they are not removed before any 
stored charge has been discharged at the end of the test. Failure to comply with these 
instructions might result in an arc being produced, which might be dangerous for the 
instrument and the operator.

Having selected inductive mode, connect all four leads to the item being tested and 
press the Test button.

DLRO will check that all four leads are in adequate contact with the test piece and 
will then apply a test current and try to find the correct range. The display will show 
1 - - followed by 1 - - - -. During this time you may see the current range increasing 
or decreasing. 

On the DLRO10 the “I” lamp will flash and the message “Current Flowing” will 
appear. 

On the DLRO10X display “Precharge” will appear briefly followed by arrows moving 
left to right, the test current will auto-range up and then “Measuring” will be 
displayed on the screen. Once readings have stabilised, the operator can then press 
the test button to stop the test. The instrument will then auto-range down and show 
“Discharging” as the energy stored in the inductive load is discharged and finally 
“Complete” will be displayed, at this point it should be safe to disconnect the “C” 
Leads.

The time required for a stable reading may vary from a few seconds up to several 
minutes depending on the inductance and resistance of the test sample. There is no 
time limit for the inductive mode test, which will continue until the operator presses 
the Test button.

When the test is terminated, the “I” lamp will remain illuminated on the DLRO10 
When these messages are extinguished it should be safe to disconnect the “C” leads. 

However, the discharge indicator is an active electronic device and should not be 
relied upon. You should observe your organisation’s operating procedures.

Содержание DLRO-10

Страница 1: ...DLRO10 and DLRO 10X Digital Low Resistance Ohmmeter User Guide User Guide DLRO10 and DLRO10X Digital Low Resistance Ohmmeter...

Страница 2: ...ltage Warning 8 4 4 Discharge Current Warning 8 4 5 The Menu 8 4 6 Test Menu 9 4 7 Options Menu 9 4 7 1 Retrieve 9 4 7 2 Download 10 4 8 Passbands 10 4 9 Set Clock 11 4 10 Delete Data 11 4 11 Storage...

Страница 3: ...8 7 3 1 Standard Charging 18 7 3 2 WARNING 18 8 SPECIFICATIONS 19 9 TROUBLE SHOOTING 21 10 ACCESSORIES 22 11 REPAIR AND WARRANTY 24 11 1 Instrument Repair and Spare Parts 24 11 2 Approved Repair Compa...

Страница 4: ...to achieve this For testing large inductive resistances the current carrying leads should be connected securely to the item under test before starting the test It is not recommended that duplex handsp...

Страница 5: ...ectrical outlets CAT IV Measurement category IV Equipment connected between the origin of the low voltage mains supply and the distribution panel 1 1 Symbols used on the instrument G Caution Refer to...

Страница 6: ...at the back and held closed by magnetic catches covers the top panel This lid may be removed if required by opening fully and snapping out of its hinges Both instruments are protected against accident...

Страница 7: ...he lamp will not operate if the instrument is turned off The I lamp will light if a current greater than 1 mA is still flowing after a test is completed This suggests that an inductive load has been t...

Страница 8: ...measurement cannot be made if any of these messages is visible on the display 4 3 3 External Voltage Warning If external voltage is applied to the terminals a message EXT VOLTS will flash on the displ...

Страница 9: ...mode 4 7 Options Menu The Options menu has five options that are not related to each other as shown below There is a quick access menu that can be accessed using the numbers keys as follows 0 Retriev...

Страница 10: ...on allows you to set upper and lower limits between which the test result average must fall if it is to be signalled as a Pass GREEN Readings outside these limits will be signalled as a Fail RED Set t...

Страница 11: ...o clear DLRO10X s memory of stored data In case you have selected this Option by accident you will be asked to confirm that you wish to delete the data the default is NO Change this to YES and press E...

Страница 12: ...is automatically limited to 1A When Battery level is low 10 the maximum test current is automatically reduced by one level and will be further reduced when a test has to be abandoned because the batt...

Страница 13: ...he test number to 1 4 15 Keypad Operation The 12 key keypad is similar to those used on mobile telephones in that each key is capable of producing one of several characters depending on the number of...

Страница 14: ...ts on loads that exhibit an inductive component using the test mode indicated by the inductor symbol on the DLRO10 or by using the inductive mode on the DLRO10X When measuring inductive loads it is es...

Страница 15: ...as shown in the next section Testing Using Duplex Handspikes or Individual Leads One of the DH4 handspikes is fitted with two lamps marked L1 and L2 and an extra lead This lead should be plugged into...

Страница 16: ...two measurements Both measurements are displayed by DLRO 10X along with the average while the DLRO 10 displays the average only If the combined resistance of the current leads and test sample is grea...

Страница 17: ...e retaining clips and pull the top of the clip away from the instrument body The module now unplugs from the base of the instrument Plug in the charger or connect to a 12 volt vehicle type battery usi...

Страница 18: ...Charging The battery module is charging at its standard rate Fast Battery state flashing during charging indicates temperature fault 7 3 2 WARNING Connecting to greater than 15 volts can cause perman...

Страница 19: ...e and Safety LED Indication Measurement Mode Manual Auto Continuous Inductive Manual Auto Continuous Inductive Unidirectional Control Fully Automatic Fully Automatic with manual override of maximum cu...

Страница 20: ...to 122 F at reduced accuracy Storage 30 C to 70 C 22 F to 158 F Calibration 20 C 68 F Coefficient 0 01 per C from 5 C to 40 C 0 006 per F from 41 F to 104 F Standard charging 0 C to 40 C 32 F to 104...

Страница 21: ...RR 114 Checksum Failure in EEPROM Calibration constants have been lost The DLRO will continue to work but the accuracy can now not be guaranteed At switch on the display will show the software version...

Страница 22: ...3 172 DLRO10LPU UK Mains Line Power Unit UK plug 1003 093 DLRO10LPU US Mains Line Power Unit US plug 1003 171 DLRO10LPU line power unit The DLRO10 and DLRO10X may be also powered from an optional main...

Страница 23: ...06 18 9 m 30 ft 242006 30 Single Leads Single handspikes 2 for potential measurement 2 m 7 ft 242021 7 5 5 m 18 ft 242021 18 9 m 30 ft 242021 30 Current clips 2 for current connections 2 m 7 ft 242041...

Страница 24: ...r Repair If returning an instrument to the manufacturer for repair it should be sent freight pre paid to the appropriate address A copy of the Invoice and of the packing note should be sent simultaneo...

Страница 25: ...cell is located behind the display Dispose of in accordance with local regulations The lithium ion is classified as an industrial battery For disposal in the UK contact Megger Ltd For disposal of batt...

Страница 26: ...USA T 800 723 2861 USA only T 1 214 333 3201 F 1 214 331 7399 E USsales megger com Megger AB Rinkebyv gen 19 Box 724 SE 182 17 DANDERYD T 46 08 510 195 00 E seinfo megger com Megger USA Fort Collins...

Отзывы: