www.megger.com
DLRO10 and DLRO10X
19
8. SPECIFICATIONS
Ranges
Full Scale
Resolution
Accuracy
Full Scale Volts
Test Current
Resistance
Induction
Resistance
Induction
1,9999 mΩ
0,1 μΩ
±0,2%±0,2 μΩ
20 mV
10 A
1 A
19,999 mΩ
1 μΩ
±0,2%±2 μΩ
20 mV
20 mV
1 A
1 A
199,99 mΩ
10 μΩ
±0,2%±20 μΩ
20 mV
200 mV
100 mA
100 mA
1,9999 Ω
100 μΩ
±0,2%±0,2 mΩ
20 mV
200 mV
10 mA
10 mA
19,999 Ω
1 mΩ
±0,2%±2 mΩ
20 mV
200 mV
1 mA
1 mA
199,99 Ω
10 mΩ
±0,2%±20 mΩ
20 mV
200 mV
100 μA
100 μA
1999,9 Ω
100 mΩ
±0.2%±0.2 Ω
200 mV
200 mV
100 μA
DLRO 10
DLRO 10X
Display
Measurement
4 1/2 digit seven segment LED
Large colour backlit
Range and Safety
LED Indication
Measurement
Mode
Manual, Auto, Continuous, Inductive
Manual, Auto, Continuous, Inductive,
Unidirectional.
Control
Fully Automatic
Fully Automatic with manual override of
maximum current
Speed
<3s for forward and reverse current and to display average
Test Method
Single cycle reversing d.c. ratiometric measurements and average result calculated.
Test Current Accuracy
10%
Output Current Stability
<10 ppm per second @ 10 mA
<100 ppm per second at higher currents
Maximum Lead Resistance
100 mΩ total for 10A operation irrespective of battery condition.
Voltmeter Input Impedance
> 200 kΩ
Hum Rejection
Less than 1% ± 20 digits additional error with 100 mV peak 50/60 Hz. on the potential leads.
Warning will show if hum or noise exceeds this level.