M
AVTMBITE 3 Rev 8 April 2014
18
Step Two: Turning on the BITE 3 and Connecting the Lead Set.
The BITE 3 is a PC-based instrument running Windows CE. It will take about
30 seconds to boot up and be ready to take measurements.
1.
Turn on the BITE 3 by pressing the on/off (O | I) button. The back light
should stay lit.
2.
Connect the lead set and, if used, the CT to the BITE 3.
3.
a)
Configure a new site and string.
Configure I Strings I New
En ter New String In fo rmatio n
Site:
String name:
Baseline:
1.00
m
Ω
Warning:
15
▲
▼
%
Fail:
30
▲
▼
%
Change:
20
▲
▼
%
Deviation:
0
▲
▼
%
-or-
b)
Select a site/string to be tested (See Chapter 3 for the flowchart of the
menus.)
Measure I New Test
Select a String
Site:
I
▲
▼
String:
▲
▼
Measure I New Test
Select and Continue
: a String
Close
Site:
I
▲
▼
String:
▲
▼
Содержание BITE 3