Procedures for Testing with the DX Host
116 Baker DX 71-030 User Guide EN V10
www.megger.com
Low-voltage Armature or Coil Test Procedures
1.
Remove the low-voltage RLC test leads and set them aside.
2.
Connect the low-voltage armature test probes to the RLC test leads connector. (The
standard low-voltage RLC test leads 1 and 2 can be used to test single coils.)
3.
Check the
Status Bar
to ensure that the
Active Folder
and
Active Record
fields
display the intended target destinations for your test data.
Figure 129. Low-voltage armature or coil test start screen.
4.
Set up the start screen start screen as shown above by touching the
RLC Tests
mode
icon in the
Mode Menu
then selecting the
Armature Coil Mode
icon in the
Mode
Submenu
.
5.
Select the coil or winding type to test as shown in the example below.
Figure 130. Selecting coil type to test—single coil selected.