Electromagnetic compatibility (EMC)
WARNING
The Invia Liberty pump should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the Invia Liberty pump should be observed to verify normal
operation in the configuration in which it will be used.
WARNING
Invia Liberty is a medical device that requires special safety precautions and must be installed and
placed in operation in accordance with the attached EMC information.
WARNING
Do not use other accessories than those specified or sold by the manufacturer as replacement parts
for internal components as it may result in increased emissions or decreased immunity of the Invia
Liberty pump.
WARNING
EMC
HF (high-frequency) surgical equipment, radio networks or the like can influence the operation of the
device and may not be operated in combination with the Invia Liberty pump.
Invia Liberty is EMC-tested in conformity with the requirements of IEC
60601-1-2:2007 and IEC 60601-1-2:2014 4th Edition according clause 7
and 8.9.
Electromagnetic compatibility ( EMC, IEC 60601-1-2:2007 3rd Edition and
IEC 60601-1-2:2014 4th Edition, Table 1 )
Electromagnetic emissions
The Invia Liberty pump is intended for use in the electromagnetic environment
specified below. The customer or the user of the Invia Liberty pump should assure that it is used in
such environment.
Emission tests
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The Invia Liberty pump uses RF energy only for it’s in-
ternal function. Therefore, its RF emissions are very low
and are not likely to cause any interference with nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The Invia Liberty pump is suitable for use in all establish-
ments, including domestic establishments and those
directly connected to the public low-voltage power sup-
ply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies