21
EN
Detailed pressure mode allows to select pressure level from
– 40 to – 200 mmHg in increments of 5mmHg (5kPa to 27kPa
in increments of 1kPa)
1.
To enter the Administrative mode, unlock the display,
press and hold [
], and then press [
].
2.
Press the soft buttons [
+
] at the same time to enter
the Settings menu.
3.
To switch pressure modes, scroll down with [
] and select
Pressure set-up.
4.
Press "OK" [
] to select.
5.
To change between Standard and Detailed pressure modes, press
Selection buttons [
] or [
] to choose Standard or Detailed
mode.
6.
Press "OK" [
] to confirm.
7.
Refer to Change pressure level instructions to select pressure values.
Change pressure set-up mode
Select air leakage volume
High air leakage mode is recommended when X-large dres-
sing is used.
Default air leakage volume is Standard. If the vacuum on the wound
is < 75% of the set pressure level or the air flow is > 1500ml/min,
leakage indicator is "full" and flashes, indicating there is a big air leak
in the system. The air leakage alarm will go off within 2 minutes if the
air flow is above the defined level or after 1 minute if the set pressure
is not maintained. Follow the instructions shown on the display, or
see the Alarms chapter.
For wounds with air leakage volume between 1500ml/min and
2200ml/min, High air leakage volume mode is available. If the vacuum
on the wound is < 75% of the set pressure level or the air flow is >
2200ml/min value, leakage indicator is "full" and flashes, indicating
there is a big air leak in the system. The air leakage alarm will go off
within 2 minutes if the air flow is above the defined level (2200ml/min)
or after 1 minute if the set pressure is not maintained. There will be
no leakage alarm during the first 5 minutes which allows establishing
the set pressure on a wound with high volume without an alarm.
➡