23
EN
Change therapy mode
1.
To enter the Administrative mode, unlock the display,
press and hold [
], and then press [
].
2.
To change the therapy mode, press
"Standby"
[
3 seconds ] to put the Invia Liberty pump into the Standby
mode.
Constant mode – C
The default therapy mode is Constant.
3.
To change from Constant mode to Intermittent mode,
press [
]
“Change to Intermittent”
and then press
“On”
[
].
Intermittent mode – I
The default times for Intermittent mode are 5 minutes on and
2 minutes off.
4.
To change from Intermittent mode to Constant mode, press
[ ]
“Change to Constant”
and press
“On”
[
].
The settings possible to change are, Unit pressure, Air leakage volume, Constant and Intermittent modes,
intermittent times, language and time zone. Pump number, pump run-time and pump version can only be
viewed but not changed.
The settings can only be changed when the pump is in the Administrative mode.
1 minute after the last button has been pressed,
the Invia Liberty pump switches into Patient mode automati-
cally and the display is locked.
1 minute after the last button has been pressed,
the backlight turns off. When pump alarms or any button is
pressed the backlight will illuminate.
➡