Maintenance
Repairs
28
GA 5752 2806 GB 10
7
Defect
Source of malfunction
Corrective actions
Oversuction
Gas-jet pump oversuction
Fit a hydrophobic filter or overflow
protection device (before doing so,
clean or replace the gas-jet pump).
Tab. 5:
Corrective actions
7.4
Repairs
The following may require repairs from the manufacturer or an authorized service partner:
• Liquid has penetrated the device.
• The performance has significantly decreased.
• Inexplicable notifications appear.
• Abnormal noises occur.
• Functional faults cannot be rectified according to the measures in chapter Malfunctions and
troubleshooting [
page 27].
If defects are detected the product may not be used any longer.
Make a note of the deficiencies and the REF number on the data plate and inform the responsible
ATMOS Service.
Observe the information in chapter Sending in the device [
page 28].
7.5
Service hotline:
+49 7653 689-0
7.6
Type plate position
1
Fig. 7: Type plate position
Position of the type plate (1) on the product.
7.7
Sending in the device
Remove and properly dispose of consumables.
Clean and disinfect the product and accessories according to the operating instructions.
Place used accessories with the product.