Introduction
Versions of FINA fine regulator O2 / AIR
12
GA 5752 2806 GB 10
1
Application time:
For continuous operation; in practice short-term use on the patient
(< 30 days)
Application site:
The application site is the clinical environment and doctor’s
practices. The application of the product may only be performed by
medically trained and introduced staff.
Contraindications:
The FINA FRV O2 and AIR may not be used for the following
purposes:
• Outside the medical sector
• In MR areas
• In the home care sector
• Being operated directly by the patient
• If not reducing the vacuum of the central gas supply poses a
danger for the patient (e.g. in drainage, in paediatrics and neo-
natology). Usage in combination with disposable thoracic drain-
age systems with integrated vacuum regulation is excepted.
• For vacuum extraction
• For the aspiration of flammable or explosive liquids
• For the aspiration of smoke that is generated during HF and
laser surgery without the connection of an intermediate smoke
filter
• For the ventilation of patients
• With central gas supply systems with other supply pressures
than 500 kPa +/-10 %
The product is:
Not active
Sterility:
No sterile product
Single-use product /
reprocessing:
The device and parts of the accessories are reusable. For
information on reprocessing, cleaning and disinfection please see
the operating instructions.
1.5.4
Versions of FINA fine regulator O2 / AIR
The connection of the tapping unit to oxygen or compressed air depends on the model being
used:
•
Version A:
Tapping unit with integrated gas pin. The FINA fine regulator is fitted directly to the
terminal unit.
•
Version B:
Tapping unit with rail clamp and NIST connection. The FINA fine regulator is
designed for mounting to an equipment rail 25 x 10 mm and is supplied via a NIST connection
with oxygen or compressed air from a terminal unit connected using a connection tube with
probe.
Products and accessories are only permitted with the ISO colour coding. In Germany, Austria and
Switzerland, products with neutral colour coding are also permitted.
NOTE
The products are supplied with ISO coding. The scope of delivery includes a label
for neutral colour coding.