Maintenance
General
GA 5752 2806 GB 10
27
7
7
Maintenance
7.1
General
Maintenance, repairs and period tests may only be carried out by persons who have the
appropriate technical knowledge and are familiar with the product. To carry out these measures
the person must have the necessary test devices and original spare parts.
ATMOS recommends: Work should be carried out by an authorized ATMOS service partner. This
ensures that repairs and testing are carried out professionally, original spare parts are used and
warranty claims remain unaffected.
DANGER!
Health hazard!
The product is used in the treatment of patients. The product or some of its
components may be contaminated.
Clean and disinfect the product before maintenance and repair. Repair work may
be performed by personnel authorized by ATMOS.
7.2
Period tests
Observe the country specific requirements regarding period tests.
7.3
Malfunctions and troubleshooting
DANGER!
Defective product!
Using incorrect spare parts and accessories can cause injuries or equipment
failure.
Only use original accessories or spare parts.
Defect
Source of malfunction
Corrective actions
With the regulation
spindle open as far as it
will go, the maximum
volume flow is not
achieved
The pressure from the gas
supply system is not high
enough
Check the pressure of the gas
supply system
The sinter filter within the gas
pin may be blocked
Have the sinter filter in the plug
replaced
Fine regulator does not
work properly
Pressure fluctuations in the
gas supply system / pressure
is too low or too high
Contact the technical service
Unknown
Contact the technical service
Probe does not fit into the
terminal unit
Wrong terminal unit or gas pin
selected
Check the designation of the
terminal unit
Fine regulator is leaking
Seal is missing or defective
Have fine regulator checked
There is a leak in the housing
Accessories are not tightened Ensure that accessories are
properly seated
Unsuitable connectors
Observe the interfaces