Cleaning and disinfection
Disinfection procedure
26
GA 5752 2806 GB 10
6
6.3.3
Disinfection procedure
After each cleaning process, wipe or spray disinfect the product in accordance with the
instructions of the disinfectant manufacturer.
Ensure that the product is free of disinfectant residue.
Perform visual and functional inspections.
6.4
Special safety notes
DANGER!
Tension cracks!
Various components in the tapping unit are made of plastic materials. Solvents and
some disinfectants and some cleaning agents can soften plastic or cause tension
fissures. The gas pressure may cause the flowmeter viewing tube to burst.
Never use detergents that contain alcohol. Observe cleaning and disinfection
standards.
CAUTION!
Property damage!
Using non-colour-fast drapes can cause discolouration of surfaces.
Only use colour-fast drapes.