3
INTRODuCTION
•
use of non original MECTRON inserts
used in accordance with the settings
designed and tested on the original
MECTRON inserts. Correct use of the
settings is only guaranteed with original
MECTRON inserts;
•
Shortage of stock material (handpiece,
inserts, wrenches) to be used in case of a
standstill due to faults or problems;
•
Incorrect/omitted maintenance
compared to what is stated in
of this manual;
• Breach of the requirements and the
information contained in
of this manual;
• Breach of the requirements and the
information contained in
of this manual;
• Unauthorised repairs in accordance with
this manual.
1.4
Safety Requirements
WARNING
:
Contraindications.
Do not use the multipiezo on patients who
carry heart stimulators (Pace-makers) or
other implantable electronic devices. This
precaution also applies to the operator.
WARNING
:
Contraindications.
Do not
perform scaling treatments without water
spray in order to avoid the insert over-
heating which may cause damages to the
tooth. Treatments with no water spray can be
carried out only with “Dry Work” inserts which
do not have the water passage.
CAUTION
:
Contraindications. Ultrasonic
scaler.
Do not perform treatments on metal
or ceramic/porcelain prosthetic artifacts
(unless otherwise specified). The ultrasonic
vibration could cause decementation/
loosening of such artifacts.
WARNING
:
Contraindications.
Interference from other equipment.
An electrosurgical scalpel or other
electrosurgical units positioned near the
multipiezo device may interfere with its
correct operation.
WARNING
:
Contraindications.
Interference with other equipment.
Even if compliant with standard IEC 60601-
1-2, multipiezo may interfere with other
devices in the vicinity. The user should take
whatever steps are necessary to eliminate
or reduce the source of interference. Install
the Multipiezo at safety distance from life-
support systems. multipiezo must not be
used adjacent or stacked with other electrical
equipments. If adjacent or stacked use of
the device is necessary, normal operation
of the equipment and multipiezo, in the
configuration in which they will be used,
should be verified prior their use.
WARNING
:
Risk of explosion.
The
device cannot operate in environments where
the atmosphere is saturated with flammable
gases (aesthetic mixtures, oxygen, etc.).