multipiezo
Symbol on keyboard
Possible cause
Solution
Peristaltic pump malfunction
Check that there are no
impediments for the rotation
of the pump.
Check that the pump and
the two pipes are correctly
installed.
The device has been
switched off and on again
without waiting 5 seconds
Turn off and wait for 5
seconds before switching on
the device again.
Faults on the electrical
network or excessive
electrostatic discharges or
internal faults
Turn off and wait 5 seconds
before switching the device
on again
If the warning persists,
contact PIEZOSuRGERY
®
Inc.
Customer Service.
Incorrect start up procedure:
the device was started with
the pedal pressed
Check that the pedal is
not pressed down. If the
problem persists, disconnect
the pedal and potentially
contact PIEZOSuRGERY
®
Inc.
Customer Service.
Table 6 –
Error Messages
NOTE:
For diagnostic warnings not included
in this list, contact PIEZOSuRGERY® Inc.
Customer Service.