54
multipiezo
11.1
Electromagnetic Compatibility IEC/EN 60601-1-2
WARNING:
Contraindications.
Interference with other equipment
Even if compliant with standard IEC/EN
60601-1-2, multipiezo may interfere with
other devices in the vicinity. multipiezo
must not be used near or stacked with
other devices. Install multipiezo far from
subsistence equipment. However, if this
is necessary, the correct operation of the
equipment in that configuration, and of all
equipment, must be checked and monitored
before starting the intervention.
WARNING:
Portable and mobile radio
communication equipment may influence the
correct operation of the device.
WARNING:
Contraindications.
Interference from other equipment
An electrosurgical scalpel or other
electrosurgical units positioned near the
multipiezo device may interfere with the
correct operation of the device itself.
WARNING:
The device requires particular
EMC precautions and must be installed
and put into service according to the EMC
information provided in this chapter.
WARNING:
Only use original Mectron
accessories and spare parts. The use of
cables and accessories not supplied by
Mectron might negatively affect the EMC
performances.
11.1.1
Guide and Manufacturer’s Declaration - Electromagnetic
Emissions
multipiezo is designed to operate in the electromagnetic environment specified below. The
purchaser or user of multipiezo should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment Guidance
RF Emissions
CISPR 11
Group 1
multipiezo uses RF energy only for its internal
operation. Therefore, its RF emissions are
very low and probably do not cause any
interference with nearby electronic devices.
RF Emissions
CISPR 11
Class B
multipiezo is suitable for use in all buildings,
including domestic buildings, and those
directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Emissions of fluctuations
voltage/flicker
IEC 61000-3-3
Compliant