
33
CLEANING AND STERILISATION
Dry the parts using a dry, non-abrasive,
low-lint cloth.
2
CAUTION
: Do not sterilise the parts
in question. They may stop working
and cause damage to people and/or
property.
WARNING
: Always switch off
the device using the O/I switch and
disconnect it from the power mains
before carrying out cleaning tasks.
WARNING
: The device and its
parts that cannot be sterilised are not
protected against the penetration of
liquids. Do not spray liquids directly on
the surface of the device or its parts that
cannot be sterilised.
CAUTION
: Do not use running water to
clean the parts in question.
CAUTION
: Do not soak these parts in
liquid and/or various kinds of solutions.
CAUTION
: If disinfection is required,
water-based disinfectant solutions must be
used with neutral pH (pH7). Alcohol-based
disinfectant solutions and hydrogen peroxide
are contraindicated, because they can
fade the colour and/or damage the plastic
materials. This is also valid for chemical
products such as acetone and alcohol.
Содержание Multipiezo
Страница 2: ...M O N G F J I A L D E C B K O RING 6 5X2 NBR x6 H P x 3 1 6 2 5 7 3 4 ...
Страница 6: ...IV multipiezo PAGE LEFT INTENTIONALLY BLANK ...
Страница 74: ......
Страница 75: ......