Operating Manual
Operation instructions
Page13-11
DANGER
Risk of injury!
The sensor element is continuously exposed to the box gas and
therefore is at risk from contamination.
►
Ensure that all local and national safety guidelines are followed
when handling potentially contaminated materials.
!
NOTE!
MBRAUN
recommends a maintenance cleaning procedure every 3
months.
CAUTION
Risk of damage!
When cleaning the sensors it is important that contamination from
the ambient atmosphere is prevented. Therefore, MBRAUN
recommends that the box parameters are set to a pressure of
b1.0 and +5.0 mbar (
see parameters chapter
) and that the
circulation mode is switched OFF.For Glove Box systems that must
remain at a negative pressure, special procedures may be required.
Contact
MBRAUN
service for advice.
This routine maintenance consists in cleaning the platinum winding of the
MB MO-SE-1
and
moistening it with phosphorous acid H3PO4. The following aids are required for
disassembling and maintaining the
MB MO-SE-1.
Tool for disassembly (screwdriver)
Soft, absorbent, lint free cloth (cotton)
Small quantity of phosphorous acid (H3PO4).
Protective clothing, including gloves and goggles
One dummy plug for the open circulation piping (DN40)
Содержание EasyLab
Страница 1: ...INERTGAS TECHNOLOGY Operating Manual EasyLab...
Страница 47: ...Operating Manual Operation instructions Page 6 7 Press to decrease the value Press to increase the value...
Страница 91: ...Operating Manual Operation instructions Page13 13...
Страница 92: ...Operating Manual Operation instructions Page13 14...
Страница 109: ...Operating Manual Operation instructions Page18 1 Chapter18 18 1 Common spare parts...