![MBRAUN EasyLab Скачать руководство пользователя страница 87](http://html1.mh-extra.com/html/mbraun/easylab/easylab_operating-manual_1752773087.webp)
Operating Manual
Operation instructions
Page13-9
Mechanical
Length over all:
205 mm, height 80 mm, depth 58
mm
Sensor-part:
length 42 mm, diameter 14 mm
Flange:
NW 40 KF
Weight:
0.7 kg
Electrical
Supply voltage:
24 VDC ±10%
Environment
Ambient temperature:
+15 to +27 °C
Pressure:
800 to 1200 mbar
(Differential pressure sensor to
electronics max.
≈
100 mbar)
Measuring
Range:
0 - 500 ppm moisture
Sensitivity:
20 mV / ppm
Response time (0 - 90 %):
approximately 120 sec. (0 -
90 %)
Warm-up time:
10 min (for < 10 ppm approx. 6
hr)
Accuracy
1)
:
High precision range
better than 5 % of value
(0 - 10 ppm):
Wide range (10 - 100
ppm):
better than 20 % of value
Drift at 10 ppm
10% / year
Sensor life
2)
:
ca. 5 years
1)
without interfering gases like NH
3
2)
with regular maintenance
13.4.4 Connection
The connection for the Moisture Sensor is made with an RJ45 (8-pole) Socket Connector. The
pin layout is shown in the table below.