Operating Manual
Operation instructions
Page13-6
13.3.2 Construction
The
MB-OX-SE1
consists of the sensor and the special electronics separated by a gas-
tight NW40 clamp flange. The sensor is protected against physical damage by a protective
cage. The sensor leads are connected to the electronics by a vacuum-tight feed-through.
The electronics are contained in an airtight box mounted directly to the back of the NW40
flange.
13.3.3 Technical Data
1. In clean argon-atmosphere, without interfering gases like H
2
O or CO
2
2. In absence of reactive gases (contact MBRAUN Service for further advice)
13.3.4 Connection
The connection for the Oxygen Sensor is made with an RJ45 (8-pole) Socket Connector. The
pin layout is shown in the table below.
Pin-No.
Contact
1
Supply Ground
2
Switching ON/OFF 24
V
3
Signal Ground
4
Live bit (O2)
5
Not Connected
6
Signal 0 - +10 V
7
24 V
8
Supply Ground
Installation
The oxygen probe is mounted on an appropriate vacuum-tight NW40 flange by means of a
centring ring and a clamp. The plug connection to the control unit should not be made before
the whole box-system has been purged sufficiently with inert gas. The operation of the probe
Содержание EasyLab
Страница 1: ...INERTGAS TECHNOLOGY Operating Manual EasyLab...
Страница 47: ...Operating Manual Operation instructions Page 6 7 Press to decrease the value Press to increase the value...
Страница 91: ...Operating Manual Operation instructions Page13 13...
Страница 92: ...Operating Manual Operation instructions Page13 14...
Страница 109: ...Operating Manual Operation instructions Page18 1 Chapter18 18 1 Common spare parts...