Operation instructions
Page16-4
Type of System
Quarterly
Annuall
Glove Box
•
Check the Omega
sealing for the windows
•
Check the gloves and
glove ports
•
Check the magnetic valves
•
Complete leakage test
•
Function test
•
Check and if necessary
replace the sealing for
the windows
•
Check the Omega sealing
of the windows
•
Check the gloves and
glove ports
•
Check the
illuminating
equipment
•
Check and if
necessary replace the
dust filters
•
Check and if necessary
replace the magnetic
valves
Gas Purification System
•
Check the magnetic valves
•
Check the blower
•
Check the vacuum pump
•
Complete leakage test
•
Function Test
•
Check the vacuum pump
•
Check and if necessary
replace the circulation
blower
•
Check and if
necessary replace the
filter medium
•
Dismantle pipe-work
and clean it. Replace all
Viton seals
•
Check and if
necessary replace the
valve seals
•
Check the cooling system
•
Check the cooling fluid
Analyzers
•
Check the sensors
•
Check the flow rate meter
•
Complete leakage test
•
Check and if
necessary replace
sensors
•
Check the vacuum pump
•
Leak test piping
•
Complete leakage test
•
Check calibration
Содержание EasyLab
Страница 1: ...INERTGAS TECHNOLOGY Operating Manual EasyLab...
Страница 47: ...Operating Manual Operation instructions Page 6 7 Press to decrease the value Press to increase the value...
Страница 91: ...Operating Manual Operation instructions Page13 13...
Страница 92: ...Operating Manual Operation instructions Page13 14...
Страница 109: ...Operating Manual Operation instructions Page18 1 Chapter18 18 1 Common spare parts...