Cleaning / Disinfection
7.0 Cleaning / Disinfection
7.2
Disinfection
7.2.1
Follow the general safety instructions as stated in section 2.0.
7.2.2
Disinfection procedures
I. Only use wipe disinfectants; do not use any spray disinfectants.
II. The procedures adopted must comply with the applicable statutory provisions and guidelines.
The operator must define hygiene guidelines and the corresponding safety measures for the
disinfection procedures used.
III. If the system is contaminated with infectious or potentially infectious material (e.g. blood,
bodily fluids, excrement), the relevant surfaces must be disinfected immediately.
IV. Surfaces should be routinely disinfected in accordance with the hygiene/disinfection system
established by the operator.
V. The combined use of different disinfectants and/or disinfectants and cleaning agents may
impair product performance or damage system surfaces.
VI. Always consider the material compatibility (e.g. stainless steel, aluminium, plastics, lead acrylic
glass, lead glass, PVC, silicon, natural or synthetic rubber, as well as chromed, painted or coated
surfaces).
VII. If unsure, apply the disinfectant on a small test area.
CAUTION – Risk of explosion.
If disinfectants are used that form explosive gas mixtures, they pose a risk to life
and health because of the risk of explosion.
!
Содержание Portegra2
Страница 2: ...2 ...
Страница 11: ...11 EN DE Operation ...
Страница 16: ...16 Positioning of extension arm spring arm Fig 9b Ceiling Column Ceiling Column x Fig 9a Fig 10 1 3 2 4 ...
Страница 20: ...20 Servicing 1 1 1 2 Fig 11 Fig 13 Fig 12 3 3 Twin column 360 column 4 5 5 6 ...
Страница 23: ...23 EN DE Notes ...
Страница 24: ...24 Notes ...
Страница 26: ...26 ...
Страница 35: ...35 EN DE Bedienung ...
Страница 40: ...40 Auslegerarm Federarm Stellung Abb 9b Aufhängungsachse Aufhängungsachse x Abb 9a Abb 10 1 3 2 4 ...
Страница 44: ...44 Wartung 1 1 1 2 Abb 11 Abb 13 Abb 12 3 3 Twin Stativ 360 Stativ 4 5 5 6 ...
Страница 47: ...47 EN DE Notizen ...