Rad-87 Pulse CO-Oximeter Operator’s Manual
5-7
5
a l a r m s a n d m e s s a g e s
M e s s a g e s
The Rad-87 Pulse CO-Oximeter will indicate other data or system errors. Message condi-
tions for the Rad-87 follow:
DISPLAY
TYPE
SOLUTION
SCROLLING ZEROS
Pulse Search
Wait for found pulse. (This Search
should occur whenever a sensor is first
applied to a patient).
PULSE BAR (SIQ)
TURNS RED
Low Signal IQ
1. Rule out occlusion of blood flow.
2. Verify placement of sensor.
PERFUSION BAR (PI)
TURNS RED
Low Perfusion
1. Rule out occlusion of blood flow.
2. Attempt to warm patient.
3. Move sensor to better perfused site.
NOTE:
Masimo recommends using
an adhesive sensor whenever
low perfusion is expected or
evident.
PARAMETER/
MEASUREMENT
LABEL AND
NUMBER FLASH
Alarm Limit Exceeded
Assess /address patient condition.
Re-set alarm limits if indicated.
System Fault
Return for service.
There are several error codes, all error
codes require return of the device to an
authorized service center for repair. See
Section 9,
Service and Repair.
Defective Sensor
Replace sensor.
Unrecognized Sensor
Connect appropriate cable.
Interference Detected
Ensure that the sensor is properly
applied, and cover the sensor site with
opaque material, if required.