4-20
Rad-87 Pulse CO-Oximeter Operator’s Manual
o p e r a t i o n
4
SETUP MENU LEVEL
2
(CONTINUED)
BUTTON VOLUME
BUTTON
SETTING
Press the Enter Button
again to move to the
next menu.
5x
Default
Level 2
Use Up or Down Button to move
between settings and the Enter
Button to accept the setting and
move to the next menu screen.
OR
press the Display Button to exit
without saving the new setting
and to return to the home display
screen.
Level 1
Off
Level 3
FASTSAT
FastSat enables rapid tracking of arterial oxygen saturation changes. Arterial oxygen
saturation data is averaged using pulse oximeter averaging algorithms to smooth the
trend. When the Rad-87 is set to FastSat “On”, the averaging algorithm evaluates all the
saturation values providing an averaged saturation value that is a better representation of
the patient’s current oxygenation status. With FastSat, the averaging time is dependent on
the input signal.
BUTTON
SETTING
Press the Enter Button
again to move to the
next menu.
6x
Default
Off
Use Up or Down Button to move
between settings and the Enter
Button to accept the setting and
move to the Alarm Volume menu.
OR
press the Display Button to exit
without saving the new setting and
to return to the home display screen.
On