3-2
Rad-87 Pulse CO-Oximeter Operator’s Manual
■
IF THERE IS ANY DOUBT ABOUT THE INTEGRITY OF THE PROTECTIVE EARTH
CONDUCTOR ARRANGEMENT, OPERATE THE RAD-
87
ON INTERNAL BATTERY
POWER UNTIL THE AC POWER SUPPLY PROTECTIVE CONDUCTOR IS FULLY FUNC-
TIONAL.
■
TO ENSURE PATIENT ELECTRICAL ISOLATION, CONNECT ONLY TO OTHER
EQUIPMENT WITH ELECTRICALLY ISOLATED CIRCUITS.
■
DO NOT CONNECT TO AN ELECTRICAL OUTLET CONTROLLED BY A WALL SWITCH OR
DIMMER.
INITIAL BATTERY CHARGING
Before use, the Rad-
87
battery needs to be fully charged.
To charge the internal battery, connect the AC power cord to an AC outlet and to the Power
Entry Module located on the back of the Rad-
87
. The AC Power Indicator illuminates. The AC
Power Indicator will remain illuminated while the battery is charging. The Battery Charge Level
Indicator will not be illuminated unless the unit is operating on battery power. Once the battery is
fully charged, the device has approximately 4 hours of battery life.
INITIAL INSTALLATION
Place the Rad-87 on a stable hard flat surface near the patient. Always place the Rad-
87
on a dry surface. Maintain a minimum of 1 inch (2.54 cm) free space around the device.
Make sure that Rad-87 loudspeaker is not covered to avoid a muffled alarm sound.
The Rad-
87
should not be operated outside the following environmental conditions:
OPERATING ENVIRONMENTAL CONDITIONS
TEMPERATURE
+5°C to +40°C, +41°F to +104°F
HUMIDITY
5% to 95%, non-condensing
OPERATING ALTITUDE
500 mbar to 1060 mbar pressure
-1000 ft to 18,000 ft (-304 m to 5,486 m)
Configure the Rad-87 for your regional power line frequency (50 or 60 hz) if needed.
Configure the Rad-87 for your regional power line frequency (50 or 60 hz) if needed.
Default is 60 hz (standard for the United States). See Section 4,
Default is 60 hz (standard for the United States). See Section 4,
Operation, Setup menu
Operation, Setup menu
Level 3, Line Frequency.
Level 3, Line Frequency.
CAUTION:
CAUTION:
THE DEVICE MUST BE CONFIGURED TO MATCH YOUR LOCAL POWER
THE DEVICE MUST BE CONFIGURED TO MATCH YOUR LOCAL POWER
LINE FREQUENCY TO ALLOW FOR THE CANCELLATION OF NOISE
LINE FREQUENCY TO ALLOW FOR THE CANCELLATION OF NOISE
INTRODUCED BY FLUORESCENT LIGHTS AND OTHER SOURCES.
INTRODUCED BY FLUORESCENT LIGHTS AND OTHER SOURCES.
CAUTION:
CAUTION:
THE BATTERY SHOULD BE ADEQUATELY CHARGED TO ENSURE BACK-
THE BATTERY SHOULD BE ADEQUATELY CHARGED TO ENSURE BACK-
UP POWER IN CASE OF AC POWER DISRUPTION.
UP POWER IN CASE OF AC POWER DISRUPTION.
s e t u p
3-2
3
RAD-
87