4-38
Rad-87 Pulse CO-Oximeter Operator’s Manual
o p e r a t i o n
4
H O M E M O D E O P E R A T I O N
The Rad-87 can be placed into the Home Mode to protect unqualified users from changing the
Rad-87 alarm settings and operation. Only the following menu and front panel functions are
available: display brightness, pulse beep volume adjustment and alarm silence. Alarm volume is
at highest setting. All default and user defined default settings are locked to their current values
when home mode is selected and return to those values after a power cycle. Upon power up,
the Hnn mode will be displayed along with a 10 second display of parameters/measurements.
To turn the device off the Power Button must be depressed and held for 3 seconds. The Enter
and Up Buttons held simultaneously for 5 second will put it back into the special menu to select
a different mode.
SLEEP MODE OPERATION
The Rad-87 can be placed into the Sleep Mode to allow the device to capture normal and
abnormal patient data without triggering the alarms. This mode will blank out the LED and the
LCD Displays with the exception of the AC Power Indicator, the System Status Light, and the
Battery Level Indicator and disables the alarms even after a power cycle. However, pressing
any button illuminates the display and the System Status Light (solid yellow) for 10 seconds.
Upon power up, the SLP mode will be displayed along with a 10 second display of parameterr/
measurement settings. The Enter and Up Buttons held simultaneously for 5 seconds will put it
back into the special menu to select a different mode.