Rad-87 Pulse CO-Oximeter Operator’s Manual
4-1
4-1
4
o p e r a t i o n
I n t r o d u c t i o n
To operate the Rad-
87
system effectively, the device must be set up correctly and the
operator must:
■
Know how the Rad-
87
derives its readings (see Section 1).
■
Be familiar with its controls, components and operation.
■
Understand its status and alarm messages (see Section 5,
Alarm and Messages
and Section 6,
Troubleshooting
).
B a s i c o p e r a t i o n
GENERAL SETUP AND USE
1. Inspect the Rad-
87
case for damage.
2.
Connect a patient cable or a direct connect sensor to the Rad-
87
device. Make sure
it is a firm connection and the cable is not twisted, sliced or frayed.
3.
If utilizing a patient cable, select a sensor that is compatible with the Rad-
87
and the
patient before connecting it to the patient cable. See section 8,
Sensors and Patient
Cables
. If using a single patient adhesive or disposable sensor, check that the emitter
(red light) and the detector are properly aligned. Remove any substances that may
interfere with the transmission of light between the sensor’s light source and detec-
tor.
4. Refer to the Directions for Use of the sensor before attaching the sensor to the patient.
5.
Attach the sensor to the Patient. With a Masimo Pulse CO-Oximetry sensor, connect
the sensor to the patient cable with the logos lining up; make sure it is a firm connec-
tion.
6. Press the Power button to turn the Rad-
87
on.
7.
Verify all front-panel indicators momentarily illuminate and a tone is heard.
NOTE:
The number "0" scrolls across the screen as the system calibrates and obtains
patient data (approximately 20 seconds).
8.
Verify the front-panel display is free of alarm and system failure messages (see
Section 5,
Alarms and Messages
).
9.
Verify the LED and the LCD displays shows the following (see
Setup Menu Level 1;
Parameter/Measurement Alarm Limits - Screen 1, Parameter/Measurement Alarm
Limits - Screen 2, Parameter/Measurement Alarm Limits - Screen 3,
and
Setup
Menu Level 3;
Set Mode
located in this chapter):
■
Mode setting: Standard (Std) or Sleep (SLP) or Home (Hnn),
■
SpO
2
Low Alarm Limit and SpO
2
High Alarm Limit,
■
Pulse Rate Low Alarm Limit and Pulse Rate High Alarm Limit,