Rad-87 Pulse CO-Oximeter Operator’s Manual
5-5
5
a l a r m s a n d m e s s a g e s
ALARM SILENCE
Audible alarms may be silenced, while visual alarms remain active. The alarm silence
function is controlled by pressing the Alarm Silence Button.
The Alarm Bell and the System Status Light provide visual feedback when the Rad-87
audible alarms are silenced.
Alarm Silence function when monitoring a patient:
Power-On – Alarms active, Alarm Bell flashes red for high priority alarms and the System
Status Light is solid yellow for low priority alarms, flashes yellow for medium
priority alarms or flashes red for high priority alarms.
Press Once – Alarm suspended for 120 seconds, audible alarms are silenced.
Press Twice – Return to audible alarms active, audible alarms return.
Repeated pressing of the Alarm Silence Button will cycle though alarm silence options above.
Alarm Silence function when not monitoring a patient:
Power-On – Alarms active, Alarm Bell is not flashing and the System Status Light is solid
yellow.
Press Once – Device is silenced until it is cycled off/on or until monitoring begins.
– The Alarm Bell is not flashing and the System Status Light is solid yellow for
low priority alarms.
ALARM BELL
The Alarm Bell flashes red for high priority alarms. Pressing the Alarm Silence Button once
silences the audible alarm for 120 seconds (default) while the Alarm Bell flashes to indicate
an alarm condition. If the high priority alarm condition is resolved during the Alarm silence
interval, the Alarm Bell stops flashing. If the high priority alarm condition remains (Alarm
Bell flashing red), pressing the Alarm Silence button again activates the audible alarms and
the Alarm Bell continues to flash red. The Alarm Bell stops flashing when the high priority
alarm conditions are resolved.