Rad-87 Pulse CO-Oximeter Operator’s Manual
5-2
a l a r m s a n d m e s s a g e s
5
A l a r m i n d i c a t i o n
An alarm condition caused by an out-of-limit parameter/measurement is indicated by:
■
Audible alarm tone
■
Flashing Out-of-limit parameter/measurement label and value
■
System Status Light flashes red for high priority alarms, flashes yellow for medium
priority alarms or shows solid yellow for low priority alarms
■
Red flashing Alarm Bell for high priority alarms
A l a r m l i m i t s
CAUTION:
TO ENSURE THAT ALARM LIMITS ARE APPROPRIATE FOR THE PATIENT
BEING MONITORED, CHECK THE LIMITS EACH TIME THE RAD-87 IS USED.
When an alarm limit is exceeded, an audible alarm activates and the Alarm Bell flashes red
for high priority alarms and the System Status Light is solid yellow for low priority alarms,
flashes yellow for medium priority alarms or flashes red for high priority alarms. When a
sensor is not connected to a patient, “SEN OFF” message will show on the display. When
a sensor is not connected to its cable, “NO SEN” message will show on the display.
NOTE:
An audible alarm will accompany the visual indicators unless the Rad-87 has
been set to Interface Alarms "Off" (only SpO2 and BPM alarms muted) or to
Sleep Mode (all alarms muted).
ALARM LIMIT: USER CONFIGURABLE SETTINGS
SETTING
RANGE
SpO
2
High Limit
The SpO
2
high alarm limit can be set anywhere between 2% and 99%,
then "---" (Off), with a 1% step size. In the “----” (Off) setting, the SpO
2
High Limit alarm is disabled.
SpO
2
Low Limit
*
The SpO
2
low alarm limit can be set anywhere between 1% and 98%,
with a 1% step size.
SpCO High Limit
The SpCO high alarm limit can be set anywhere between 2% and 98%,
then "---" (Off), with a 1% step size..
SpCO Low Limit
*
The SpCO low alarm limit can be set anywhere between "---" (Off), then
1% and 97%, with a 1% step size. In the “----” (Off) setting, the SpCO
Low Limit alarm is disabled.
SpMet High Limit
The SpMet high alarm limit can be set anywhere between 1% and 99.5%,
then "---" (Off). Adjust in increments of 0.1% for values between 0.1% to
2%. Adjust in increments of 0.5% for values between 2% and 99.5%.
SpMet Low Limit
*
The SpMet low alarm limit can be set anywhere between "---" (Off),
then 0.1% and 99%. Adjust in increments of 0.1% for values between
0.1% to 2%. Adjust in increments of 0.5% for values between 2% and
99.5%.
When SpMet is placed in the “----” (Off) setting, the SpMet Low
Limit alarm is disabled.