Martin maxium & maxium Beamer Operating Instructions
V 2.0
87
14.1 Technical Specifications & Certificate
Power requirements
100–240 V ± 10%; 50/60 Hz
Line current
max. 6.3 A
Line fuse
6.3 A
Nominal frequencies
300 / 400 / 600 kHz
Power input
without HF output
40 VA
at max. power output
600 VA
Output power
maximum cutting power
400 W
maximum coagulation power
320 W
Current Type
HF Power
Crest Factor
HF Voltage
Pure Cut
5–400 W at 400 ohms
CF 1.5 max. 1,200 V
p
max. 2,400 V
pp
Blend Cut
5–320 W at 400 ohms
CF 1.9 max. 1,400 V
p
max.
2,800
V
pp
Super Blend
5–250 W at 400 ohms
CF 2.4 max. 1,400 V
p
max. 2,800 V
pp
Uro Cut Pure
5–400 W at 600 ohms
CF 1.5 max. 1,600 V
p
max.
3,200
V
pp
Uro Cut Blend
5–320 W at 600 ohms
CF 1.9 max. 1,600 V
p
max. 3,200 V
pp
Argon Cut
5–400 W at 400 ohms
CF 1.5 max. 1,200 V
p
max.
2,400
V
pp
Polyp I*
1–120 W at 150 ohms
n. a. max. 1,200 V
p
max. 2,400 V
pp
Polyp II*
1–120 W at 150 ohms
n. a. max. 1,200 V
p
max.
2,400
V
pp
Papilla I**
1–50 W at 400 ohms
n. a. max. 2,000 V
p
max. 4,000 V
pp
Papilla II*
1–100 W at 150 ohms
n. a. max. 1,500 V
p
max. 3,000 V
pp
Care Cut
1–70 W at 600 ohms
CF 1.7
max. 750 V
p
max. 1,500 V
pp
Clamp Coag
5–250 W at 250 ohms
CF 3.0 max. 1,500 V
p
max. 2,800 V
pp
Forced Clamp
5–150 W at 300 ohms
CF 4.6 max. 2,000 V
p
max. 4,000 V
pp
Contact Coag
5–250 W at 100 ohms
CF 3.3 max. 2,000 V
p
max. 4,000 V
pp
Contact Coag Auto Stop
5–250 W at 100 ohms
CF 3.3 max. 2,000 V
p
max. 4,000 V
pp
Forced Coag
5–150 W at 300 ohms
CF 5.4 max. 2,800 V
p
max. 5,000 V
pp
Uro Coag
5–150 W at 600 ohms
CF 5.7–4.4 max. 3,700 V
p
max. 6,000 V
pp
Forced Uro Coag
5–150 W at 600 ohms
CF 5.7 max. 4,500 V
p
max.
7,500
V
pp
Macro Spray
2–150 W at 600 ohms
CF 20–9.0 max. 8,000 V
p
max. 11,000 V
pp
Spray
2–150 W at 600 ohms
CF 9.0 max. 8,000 V
p
max. 11,000 V
pp
Endo Spray
1–100 W at 500 ohms
CF 18–5.4 max. 4,000 V
p
max. 5,900 V
pp
Duo Spray
2–150 W at 600 ohms
CF 9.0 max. 8,000 V
p
max. 11,000 V
pp
Care Coag
1–70 W at 100 ohms
CF 1.7
max. 250 V
p
max. 500 V
pp
Care Coag Auto Stop
1–70 W at 100 ohms
CF 1.7
max. 250 V
p
max.
500
V
pp
Desiccation
1–120 W at 100 ohms
CF 1.7
max. 350 V
p
max. 700 V
pp
Desiccation Auto Stop
1–120 W at 100 ohms
CF 1.7
max. 350 V
p
max.
700
V
pp
Argon Beam
2–150 W at 600 ohms
CF 9.0 max. 8,000 V
p
max. 11,000 V
pp
Argon Endo
1–100 W at 500 ohms
CF 18–5.4 max. 4,000 V
p
max.
5,900
V
pp