Martin maxium & maxium Beamer Operating Instructions
V 2.0
62
9.7
Laying out the Cables on the Patient
If the monopolar and bipolar methods are used jointly, be sure to maintain a minimum distance of
10 cm between the corresponding cables. Give the cables enough slack and avoid loops.
9.8
Depositing HF Instruments
HF instruments may never be placed on the patient during application pauses.
9.9 Accessories
When using accessory equipment, please observe the following:
•
The accessories used must have sufficient electric strength. Be sure to observe the corre-
sponding requirements specified in the descriptions of the various currents.
•
The rated minimum voltage of any such product must match the voltage values indicated in
section 14.1, “Technical Specifications & Certificate”.
•
Be sure to use only accessories that are in perfect working condition!
•
Only Martin accessories may be used!
9.10 Interference with Other Equipment
When using high-frequency currents for electrosurgery as intended, other electromedical equip-
ment (such as ECG monitors) or electrical equipment such as telephones or PCs can be adversely
affected due to interference caused by your maxium.
To prevent or minimize interference problems, you can take the following measures:
•
Connect the maxium to a different mains circuit (use different socket outlet).
•
Increase the distance between the maxium and any equipment showing interference symp-
toms.
•
In addition, install a commonly available mains filter on the PC or telephone system.
If the problems cannot be fixed by these measures, please contact the Martin Hotline.