Section 16: MP-205 Service Manual; 044540A-0296
16-8
Table 16-3: Reported Errors
Error #
Error Description
Recommended Action
E011
RAM error occurred
(MP-205 stops normal
operation)
Host should not continue operation
with the MP-205.
E012
ROM error occurred
(MP-205 stops normal
operation)
Host should not continue operation
with the MP-205.
E013
Last message sent by host
had a checksum error
(last message or command
was discarded)
Serial line may be defective. Check
for possible noise on the serial line
interface. Verify correctness of
checksum calculation. Resend last
message to the MP-205. If error
persists, MP-205 may be defective.
E014
Write not permitted or
command not allowed
(command discarded, normal
operation continues)
Check command description for valid
values. Resend command with
correct value.
E015
Module received a command
with a value out of range
(command discarded, normal
operation continues)
Check command description for valid
values. Resend command with
correct value.
E016
Calibration failed
(MP-205 continues with
normal operation)
Nonfatal error; host design decides
what action to take.
E017
Calibration already in
progress
(MP-205 continues with
normal operation)
Continue processing.
E018
Calibration request denied
(MP-205 continues with
normal operation)
Continue processing.
E019
Communication syntax error
on command sent by host
(command discarded; normal
operation continues)
Check command description for
correct syntax. Resend command
with correct value.
E0210
Excessive sensor current was
detected (MP-205 stops
normal operation and
continuously sends this error
message)
Notify user to check and/or replace
the sensor or cable. Reset or cycle
power to the MP-205. Retry
saturation measurement with patient.
Check for excessive noise on +5 volt
digital supply. If error persists,
MP-205 may not be functioning
properly.
16.18.3 Waveforms
Figure 16-1 through Figure 16-5 are typical waveforms as measured at various
test points (labeled TP) on the oximetry module. These waveforms are
valuable in tracing signals and locating faults. The user must use a
Nellcor
SRC-2 pulse oximeter tester. Contact Mallinckrodt Technical Services
Department if you have difficulty replicating these waveform examples.
Содержание NELLCOR NPB-4000
Страница 66: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 68: ...Section 7 Spare Parts 7 2 Figure 7 1 NPB 4000 C Top Assembly Drawing ...
Страница 70: ...Section 7 Spare Parts 7 4 Figure 7 2 NPB 4000 C Front Case Assembly Diagram Sheet 1 of 2 ...
Страница 72: ...Section 7 Spare Parts 7 6 Figure 7 3 NPB 4000 C Front Case Assembly Diagram Sheet 2 of 2 ...
Страница 74: ...Section 7 Spare Parts 7 8 Figure 7 4 NPB 4000 C Rear Case Assembly Diagram Sheet 1 of 2 ...
Страница 76: ...Section 7 Spare Parts 7 10 Figure 7 5 NPB 4000 C Rear Case Assembly Diagram Sheet 2 of 2 ...
Страница 78: ...Section 7 Spare Parts 7 12 Figure 7 6 NPB 4000 C Power Supply Heat Sink Assembly Diagram ...
Страница 80: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 96: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 114: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 140: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 178: ...Section 14 Main Color Board Digital Theory of Operation 14 38 Figure 14 14 NPB 4000C Color Motherboard Block Diagram ...
Страница 180: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 192: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 208: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 210: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 211: ...Section 17 Drawings 17 3 Figure 17 1 MP 205 PCB Schematic Sheet 1 of 2 ...
Страница 212: ...Section 17 Drawings 17 5 Figure 17 2 MP 205 PCB Schematic Sheet 2 of 2 ...