22
/
48
6.2.10
PLC Link Function
The PLC link function allows the MG80-SC to read and write the PLC’s memory and allows the PLC to operate the
MG80-SC and obtain measurement data without the need to prepare a communication program on the PLC side.
In order for the MG80-SC to read and write the memory of the PLC, the PLC must support one of the following
communication protocols.
PLC
manufacturer
Connection method
Communication protocol
Mitsubishi
Electric
Corporation
RS-232C
MC protocol 1C frame format 4
RS-232C
MC protocol 3C frame format 4
TCP/UDP
MC protocol 1E frame
TCP/UDP
MC protocol 3E frame
OMRON
Corporation
RS-232C
C-mode Commands
TCP/UDP
FINS Commands
Keyence
Corporation
RS-232C
KV Host Link Mode
TCP/UDP
Host-link
The MG80-SC provides arbitrary access to the starting addresses of three types of data area: the control flag area, the
setup parameter area, and the measurement data area.