48
6.2.4
Reset
This function resets all the measured values to zero regardless of whether the preset value is set.
Note
・
The reference point setting is also canceled when the reference point is being used. Set the reference point again.
・
If a reset is supplied while the reference point is being set, the reference point setting is canceled at that time.
6.2.5
Start
Resets the peak value.
Measured value
Change in value
Current value
Remains unchanged
Maximum value
Sets the current value
Minimum value
Sets the current value
Peak-to-peak (P-P) value
Sets the value obtained by subtracting the
minimum value (zero) from the maximum
value
6.2.6
Go/No Go Judgment (Comparator)
This function judges “go” or “no go” for the measured values in the output value mode that is set. The Go/No Go
judgment results are added to the data and output by RS-232C or Ethernet.
The comparator values set the upper and lower limits. Up to four sets of comparator values can be set, and the set of
values used can be switched partway through the measurement.
Measured value
Judgment
Judgment output
Upper limit < measured value
Upper limit NG
U
Lower limit ≤ measured value ≤ upper limit
GO
G
Measured value < lower limit
Lower limit NG
L