- 38 -
LX400+ UV LED Spot Curing System
Pin 2 INTIN2
Analog input from 1.0VDC to 4.0VDC will set the output
intensity of channel 2 from 15% to 100% in 1%
increments. The following equation can be used to
determine the output intensity based on a voltage input;
y=(85x-40)/3
where;
x is the voltage input from 1.0V to 4.0V.
y is the output intensity in percent. Value round to
nearest percent.
Pin 3 GND
Return for digital input signals.
Pin 4 ALARM (+) and Pin 16 ALARM (-)
Signal output to indicate an error has occurred. The
signal is active when any LED head alarm condition is
present. The signal is inactive at all other times.
Pin 5 SYNCOUT1 (+) and Pin 17 SYNCOUT1(-)
Signal output to indicate LED output of channel 1 status.
When the LED head is on, the signal is active. When the
LED head is off, the signal is inactive.
Pin 6 SYNCOUT2 (+) and Pin 18 SYNCOUT2(-)
Signal output to indicate LED output of channel 2 status.
When the LED head is on, the signal is active. When the
LED head is off, the signal is inactive.