![Lumen Dynamics Group OmniCure LX400+ Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/lumen-dynamics-group/omnicure-lx400/omnicure-lx400_user-manual_1947995014.webp)
- 9 -
LX400+ UV LED Spot Curing System
Head Model 385nm
Blue Light Hazard Radiance
(determined from table 5.5 of IEC 62471)
Distance Away From Head
Along Plane of Beam Axis (mm)
Distance Away From Head
Normal to Beam Axis (mm)
430
0
314
62
252
76
195
77
107
53
42
27
Head Model 385nm
Weighted Skin Thermal Hazard Irradiance
(determined from table 5.4 of IEC 62471)
Distance Away From Head
Along Plane of Beam Axis (mm)
Distance Away From Head
Normal to Beam Axis (mm)
271
0
254
24
197
43
107
50
42
27
Head Model 385nm
Un-Weighted UVA Irradiance
(determined from table 5.4 of IEC 62471)
Distance Away From Head
Along Plane of Beam Axis (mm)
Distance Away From Head
Normal to Beam Axis (mm)
261
0
254
13
197
36
107
47
42
27