LX400+ UV LED Spot Curing System
Risk Group Test Results
– Radiance Summations (watt/sr-m²)
Irradiance Summations (watt/sr-m²)
(IEC62471)
Risk Group:
365nm Head
Risk Group:
385nm Head
Risk Group:
400nm Head
Specified
Maximum
Blue Light Hazard
Radiance
Risk Group 1
Risk Group 1
Risk Group 1
1000 watt/sr-m²
(Risk Group 1)
Weighted Retina
Visible & Infrared
Thermal
Radiance
Exempt
Exempt
Exempt
622222 watt/sr-m²
(Risk Group 1)
Weighted Actinic
Eye Irradiance
Risk Group 1
Exempt
Exempt
0.003 watt/m²
(Risk Group 1)
Un-weighted UVA
Irradiance
Risk Group 1
Risk Group 1
Exempt
33 watt/m²
(Risk Group 1)
Blue Light Hazard
Irradiance
Exempt
Marginal
Exempt
Risk Group 2
400 watt/m²
(Risk Group 2)
Un-weighted Eye
IR
Exempt
Exempt
Exempt
570 watt/m²
(Risk Group 1)
Un-weighted Skin
Thermal Hazard
Irradiance
Exempt
Risk Group 1
Exempt
820 watt/m²
(Risk Group 1)
Weighted Retinal
Visible & Infrared
Irradiance Weak
Stimulus
Exempt
Exempt
Exempt
1333333 watt/m²
(Risk Group 2)