![Lumen Dynamics Group OmniCure LX400+ Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/lumen-dynamics-group/omnicure-lx400/omnicure-lx400_user-manual_1947995016.webp)
- 11 -
LX400+ UV LED Spot Curing System
Test Results
– Irradiance Summations (watt/m²)
# = complied with Risk Group 1
## = complied with Risk Group 2
(IEC62471)
Test
Results:
365nm Head
Test
Results:
385nm Head
Test
Results:
400nm Head
Exempt
Group
Specified
Max*
Weighted
Actinic UV
Skin & Eye
Irradiance
(table 5.4)
0.00241
watt/m² #
0.00092
watt/m²
0.00031
watt/m²
0.001
watt/m²
Un-weighted
UVA
Irradiance
(table 5.4)
23.44
watt/m² #
18.68
watt/m² #
3.49 watt/m²
10 watt/m²
Blue Light
Hazard
Irradiance
(table 5.4)
0.281
watt/m²
1.01
watt/m²
2.00
watt/m² ##
1.0 watt/m²
Un-weighted
Eye IR
Hazard
Irradiance
(table 5.4)
0.00016
watt/m²
0.00001
watt/m²
0.00037
watt/m²
100 watt/m²
Un-weighted
Skin Thermal
Hazard
Irradiance
(table 5.4)
2.28 watt/m²
19.14 watt/m²
#
7.56 watt/m²
8.77 watt/m²
Weighted
Retinal
Visible &
Infrared
Irradiance
Weak
Stimulus
(table 5.5)
0.0494
watt/m²
0.0043
watt/m²
0.131 watt/m²
1333333