Installation
24
Revision R09, November 2016
Fig. 8:
Position for raising and edge protection profile
4.
Place the CHM 15k in the mounting position (vertical) onto its concrete
base. As you do this, pay attention to the outer door position in relation to
the local power connection box (see Fig. 4).
5.
Use the washers and screws (supplied) to pre-assemble the CHM 15k (see
Fig. 9) so it is preliminarily fixed on its concrete base.
Fig. 9:
Fasteners (metric parts are part of the delivery, imperial sizes can be
used instead, 3/8“screw size)
1
S12 screw anchor, #12
2
±5 mm levelling screw (integrated into the device foot)
3
DIN 571-10 x 140-ZN screw or 3/8“ screw
4
ISO 7093-10.5-A2 washer
5
ISO 7093-10.5-KST/PA washer
6.
Turn the levelling screws (integrated with the support legs) until the
CHM 15k is properly aligned in a vertical direction (using a water-level
placed onto a side wall or the front face).
7.
Tighten fastening screws (nuts).
8.
Remove handle protection (edge protection profile) from top and fix it in
the pedestal for the next transport.
Concrete mounting base
Содержание CHM 15k Nimbus
Страница 6: ......
Страница 14: ......
Страница 20: ......
Страница 100: ...Revision R09 November 2016 100 Declaration of Conformity...