Safety
8
Revision R09, November 2016
refer to chapter 2.6 Safety Labelling).
2.4
Requirements on Personnel
l
The CHM 15k may only be installed and commissioned by properly trained
personnel who have received instructions on operating safety.
l
Maintenance or adjustment work on the CHM 15k may only be carried by
G. Lufft GmbH service personnel or specially trained and authorized techni-
cians of the Customer.
l
Anyone who is entrusted with work to install and commission the CHM 15k
must have completely read and understood this Manual.
l
Personnel working with the CHM 15k must not be in a state of fatigue or
under the influence of alcohol or medication or have physical impairments
of any kind that might temporarily or lastingly restrict their attention or
judgement.
2.5
Safety Notes Regarding Transportation, Installation, Com-
missioning & Cleaning
l
For handling, shipment or transportation, the CHM 15k must be duly pack-
aged and placed in the transporting position (refer to Fig. 5). Adequate
hoisting equipment and an appropriate means of transportation must be
used in all cases.
l
Once packaged, the CHM 15k must be secured and protected against acci-
dental shifting, mechanical shock or other similar impacts that may occur
inside the selected means of transportation, for example, by using tension
belts.
l
A packaged CHM 15k must not be stacked.
l
The CHM 15k requires a footprint area of 50 x 50 cm. It must be installed
and assembled in a stable and fixed position on a suitably sized concrete
foundation. The maximum permitted inclination angle of the mounting
base is 5 mm/m.
l
If the CHM 15k is not intended to be assembled immediately, it must be
protected from external influences and securely stored.
l
For installation, the following minimum prescribed clearances must be kept
betweenthe CHM 15k and:
l
Prevent irradiation from strong light sources.
l
The angle of insolation must be less or equal 15 degrees against the vertical
line. Please ask for a suiable adapter plates.
l
The assembly site must be at a great enough distance from tree plantations
or shrubbery to prevent leaves or needles restricting the cloud meter's light
outlet openings.
l
At least two persons are required for installation of the CHM 15k.
l
On completion of installation work, ensure that no safety-relevant changes
- mobile phones
2,5 m
- stationary emitters, ground stations (
100 W of output power) 25 m
- a second cloud height meter (to prevent optical interferences)
10 m