![Ludlum Measurements 3003 Скачать руководство пользователя страница 71](http://html.mh-extra.com/html/ludlum-measurements/3003/3003_user-manual_1946965071.webp)
Model 3003 and 3003i Multi-Detector Survey Meter / SCA
Section 6
Ludlum Measurements, Inc.
Page 6-1
September 2020
Software
Connecting to Software
The Model 3003 software is sent with a standard two-meter cable (A five-meter
cable can be provided if requested. However, any cable longer than two meters
may have issues with some USB hubs and computers, typically laptops.)
To connect an instrument to the computer, please connect one end of the USB
cable to the instrument first, and then the other end to the computer. Do not
connect both ends to the computer.
Please allow Windows® a moment to install the proper HID drivers for the
instrument before trying to use any software.
Note:
We recommend that you plug the USB cable into the back of your
PC that connects to your motherboard instead of a USB hub.
Note:
Some parameters may only be edited in software, such as the
backlight thresholds, COUNT Display Mode, COUNT Audio
Mode, and Setup prtct.
Follow the Lumic 2.0 software manual.