Model 3003 and 3003i Multi-Detector Survey Meter / SCA
Section 4
Ludlum Measurements, Inc.
Page 4-17
September 2020
F-F
(fixed fast at 4 seconds)
Use the PRG and ACK/RST buttons to scroll through the available parameter
options.
Response Time
- Upon entry into the user-defined Response Time, the
parameter will be selected (flashing) indicating the state of the parameter.
1 - 60 (response time in seconds from 10% of old reading to 90% of
new reading)
Use the PRG and ACK/RST buttons to scroll through the available parameter
options.
Background Audio
- Upon entry into Background Audio, the parameter will
be selected (flashing) indicating the state of the parameter.
on
(background audio is enabled)
off
(background audio is disabled)
Use the PRG and ACK/RST buttons to scroll through the available parameter
options.
Background Alarms
- Upon entry into Background Alarms, the parameter will
be selected (flashing) indicating the state of the parameter.
on
(background alarms are enabled)
off
(background alarms are disabled)
Use the PRG and ACK/RST buttons to scroll through the available parameter
options.
Sigma Audio
- Upon entry into Sigma Audio, the parameter will be selected
(flashing) indicating the state of the parameter.
on
(Sigma audio is enabled)
off
(Sigma audio is disabled)
Use the PRG and ACK/RST buttons to scroll through the available parameter
options.
Page 4 - Calibration
To begin adjusting a calibration parameter, short press the PRG or ACK/RST
to scroll through the available calibration parameters. When the appropriate
parameter is selected, short press DETECTOR to select the detector that needs
the parameter adjusted. To begin, short press LOG to begin modifying the
parameter chosen. Finally, extra-long pressing the LOG button will complete
the modification.
Содержание 3003
Страница 3: ...Model3003and3003iMulti DetectorSurveyMeter SCA User s Manual ...
Страница 7: ......