If datalogging is enabled and logging the count results was
previously requested, the results will be logged to the log buffer.
A short press of the UNITS button will switch between the Primary
and Secondary Count Units.
An extra-long press of the ACK/RESET will reset the Count Timer
and initialize for a new count.
If an alarm condition occurred during the Timed Count, a
continuous audio tone will sound, and the ALARM display indicator
will already be on. A short press of the ACK/RST button once will
acknowledge and turn off the continuous tone alarm audio. Doing
an extra-long press (pressing the button and releasing after at least a
half second) to the ACK/RST button will clear the alarm condition
and reset the Count Timer.
If other operational modes are available, a short press of the MODE
button will move to the next available operational mode.