Model 3003 and 3003i Multi-Detector Survey Meter / SCA
Section 4
Ludlum Measurements, Inc.
Page 4-9
September 2020
Extra-long-pressing the LOG button, no matter what parameter is shown, will
first return the user back to the Page Selection screen and then save and exit if
held for approximately 5 seconds in total.
When you get to a parameter that you want to change, short press the LOG
button. The display will begin flashing. This indicates the parameter can be
changed. Use the PRG or RST buttons to make the change. If numbers are
displayed on the parameter, the PRG button will increment the numbers and
RST will decrement the numbers. Otherwise, the PRG and RST button will cycle
through the available settings. When you have adjusted the parameter setting to
your desired number or setting, extra-long press the LOG button. This will save
the change and send you back up one screen to the page parameters.
Finally, when viewing a Setup Page that is the same for multiple detectors, but
each detector requires a different value, pressing the DET button until the
appropriate detector LED is illuminated will allow the user to see and change the
value for the selected detector.
For example:
Set Detector 1 Detector to: Rate Mode, Unit 1, Alarm Value 1 to 150mR/h.
Enter the Setup Menu.
Press the PRG button to advance to U-1.
Press the LOG button to enter the menu. (The first value is shown and the LO
LED should illuminate). If the LO LED is not lit, short press the DET (detector)
button to cycle through the detectors until the LO LED is lit.
Press the PRG button to get to the third of the 14 menu items. This is the
parameter setting for the “RATE/MAX ALARM 1.”
Press the LOG button to edit the value. The last digit should begin flashing.
Press the LOG button to advance to the tens digit.
Press the PRG button to increment the value to 5.
Press the LOG button to advance to the hundreds digit.
Press the PRG button to increment the value to 1.
Press the LOG button to advance to the multiplier. (All digits, multiplier, and
units will flash.)
Press the PRG button to increase the multiplier until the display shows 150 (no
decimal) and the units multiplier has advanced to mR/hr.
Press the LOG button for 2 seconds to exit.
Press the LOG button for 2 seconds to exit menu.